계측기정보 2017. 12. 2. 09:25

기기명 (Instrument) :  Documenting Process Calibrators

제작회사 (Manufacturer) :  FLUKE

형식 (Model) :  740 Series ( 741B / 743B / 744 )

세부내용 (Description) : 

741B Specification

Measurement Accuracy
-. Voltage DC
110.000 mV  0.025%+0.015% *
1.10000 V  0.025%+0.005% *
11.0000 V  0.025%+0.005% *
110.000 V  0.05%+0.005% *
300.00 V  0.05%+0.005% *
-. Voltage AC
20 to 40 Hz  2% + 10
40 to 500 Hz  0.5% + 5
500 to 1 kHz  2% + 10
1 kHz to 5 kHz  10% + 20
Ranges:  1.1000, 11.000, 110.00, 300V
-. Current DC
30.000 mA  0.01% + 0.015% *
110.00 mA  0.01% + 0.015% *
-. Resistance
11.000 Ω   0.05% + 50 mΩ
110.00 Ω   0.05% + 50 mΩ
1.1000 kΩ 0.05% + 500 mΩ
11.000 kΩ 0.1% + 10 Ω
-. Frequency
1.00 to 109.99 Hz  ; 0.05 Hz
110.0 to 1099.9 Hz   ; 0.5 Hz
1.100 to 10.999 kHz  ; 5 Hz
11.00 to 50.00 kHz   ; 50 Hz
-. Pressure
Accuracy  from 0.025% of range using any of 29 pressure modules.
   (for detailed specifications refer to pressure modules in options and accessories)
   Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
-. Note
*  (% of reading + % of full scale)


Source Accuracy
-. Voltage DC
110.000 mV  0.01%+0.005% *
1.10000 V  0.01%+0.005% *
15.0000 V  0.01%+0.005% *
-. Current DC
22.000 mA (Source)  0.01%+ 0.015% *
22.000 mA (Simulate)  0.02% + 0.03% *
-. Resistance
11.000 Ω  0.01% + 20 mΩ
110.00 Ω  0.01% + 40 mΩ
1.1000 kΩ  0.02% + 500 mΩ
11.000 kΩ  0.03% + 5Ω
-. Frequency
0.00 to 10.99 Hz  0.01 Hz
11.00 to 109.99 Hz  0.1 Hz
110.0 to 1099.9 Hz  0.1 Hz
1100 to 21999 Hz  2 Hz
22.000 to 50.000 kHz  5 Hz
-. Note
*  (% of reading + % of full scale)


RTDs and Thermocouples
-. Measure Accuracy:
10 Ω Cu (427):  2 °C
100 Ω Pt (3916):  0.3 °C
100 Ω Pt (3926):  0.3 °C
100 Ω Pt (385):  0.3 °C
200 Ω Pt (385):  0.3 °C
500 Ω Pt (385):  0.3 °C
1000 Ω Pt (385):  0.3 °C
120 Ω Ni (672):  0.3 °C
note:  For 2-and 3-wire measurement, add 0.4 °C
E:  0.3 °C
N:  0.5 °C
J:  0.3 °C
L:  0.3 °C
K:  0.3 °C
T:  0.3 °C
U:  0.3 °C
B:  0.9 °C
R:  1.0 °C
S:  0.9 °C
C:  0.6 °C
note:  Accuracy with external cold junction, for internal junction add 0.2°C
-. Source Accuracy:
10 Ω Cu (427):  1 °C
100 Ω Pt (3916):  0.1 °C
100 Ω Pt (3926):  0.1 °C
100 Ω Pt (385):  0.1 °C
200 Ω Pt (385):  0.1 °C
500 Ω Pt (385):  0.1 °C
1000 Ω Pt (385):  0.1 °C
120 Ω Ni (672):  0.1 °C
note:  For 2-and 3-wire simulation, add 0.4 °C
E:  0.2 °C
N:  0.3 °C
J:  0.2 °C
L:  0.2 °C
K:  0.3 °C
T:  0.3 °C
U:  0.3 °C
B:  0.8 °C
R:  0.9 °C
S:  0.9 °C
C:  0.6 °C
note:  Accuracy with external cold junction, for internal junction add 0.2 °C


Specifications
-. Ramp Functions
Source functions: Voltage, current, resistance, frequency, temperature
Rate: 4 steps/second
Trip detect: Continuity or voltage (continuity detection not available when sourcing current)
-. Loop Power Function
Voltage: Selectable, 24 V or 28 V
Accuracy:  5%
Maximum current: 22 mA, short circuit protected
Maximum input voltage: 30 V DC
-. Step Functions
Source Functions: Voltage, current, resistance, frequency, temperature
Manual Step: Selectable step, change with arrow buttons
Autostep: Fully programmable for function, start delay, stepvalue, time per step, repeat


Environmental Specifications
-. Operating Temperature
   -10 °C to +50 °C
   -20 °C to +50 °C *
-. Storage Temperature
   -20 °C to +60 °C
-. Dust/water resistance
   Meets IP52, IEC 529
-. Operating Altitude
   2800 m
-. Note
*  Except frequency and AC


Safety Specifications
-. Agency Approvals
   CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993


Mechanical & General Specifications
-. Size
   130 x 236 x 61 mm
-. Weight
   1.4 kg
-. Batteries
NiCd:  7.2V, 1.7 Ah
-. Battery Life
   ~8 hours typical
-. Battery Replacement
   Via snap-shut door without opening calibrator; no tools required
-. Side port connections
   Pressure module connector
   Connection for optional battery eliminator
-. Data storage capacity
   1 day of calibration results
-. 90 day specifications
   The standard specification intervals for the 741B is 1 and 2 years.
   Typical 90 day measurement and source accuracy can be estimated by
    dividing the one year "% of reading" or "%of output" specifications by 2.
   Floor specifications, expressed as "% of full scale" or
    "counts" or "ohms" remain constant.

 

첨부 (Attach)

   파일내용 (The content of file) :  Data Sheet;Ready For Anything

Fluke 741B 743B 744_Documenting Process Calibrators(Data Sheet;Ready For Anything).pdf
0.54MB

   

   사진 (Picture)  : 

posted by 엠씨고
: